Single Channel Dialyzer Machine
One button operation – With just one button press, it takes less than seven minutes to reprocess one dialyzer. Both high flux and low flux dialyzers and hemodia-filters can be reprocessed. Five step process – rinses the dialyzers, clean using liquid disinfectant (Hydrogen Peroxide and Peracetic acid), tests the membrane integrity by measuring bundle volume and pressure leak tests and finally affusion, filling the dialyzer with liquid disinfectant. Each cycle takes 350 ml (Diluted) of disinfectant per dialyzer
In built Conductivity sensor – As the chemical is prepared manually, the machine ensures the chemical has the required conductivity. The machine stops and reports with an alarm to user if the chemical isn’t premixed as per the required standards.
Dialyzer membrane integrity tests – Bundle volume is visible on the level scale for manual read. The Pressure leak test is reported on the OLED display. Disinfectant should be premixed externally to use in this machines.